TECHNICAL SPECIFICATIONS:- The Bio-Chemistry Analyzer is mainly used for the quantitative analysis of the clinical bio-chemical items; it applies the lamber- law: A=kbc=lg(i0/i). Among them:a: absorbency, k: aspirated index, b: liquid layer’s thickness, c: solution’s consistency, i0: incident light’s strength, i: transmit light’s strength. According to the above formulas, under the condition of the certain liquid layer’s thickness and incident light’s strength, as long as test the strength of transmit light which goes through the liquid, so we can calculate the consistency of the solution. Technical Parameters Method: kinetic. End-point, linearity, plus time, non linearity plus time, absorbency, sample blank, multi-standard, velocity, calculation, two-point, double wavelength etc. 7 wavelengths optic system avoiding outer disturbance. Wavelength range: 340-700nm Quality control: quality control for all tests, statistic and print quality control parameters automatically with QC curves. Every test can do high medium low three groups quality control. Program: add or delete test items easily according to the condition. Modify the test parameters. Display dynamic curves at real time. Photometric range -0. 2 - 2. 3 abs. Temperature 0degreeC-40degreeC Relative humidity 10%-90 %. Power 85-264V Dimension 33cm in length*35. 5cm in width*12. 59cm in height Communications connector RS-232 serial. Volume More than equal to 500UL Screen 240*128 dots graphic LCD. Liquid thickness 10mm. Lamp quartz-halogen lamp, 6V 10W. Measure volume 32ULL Flow cell quartz.